Sulphur dioxide

NO.5La
SO2
Performance
Measuring Range
0.5 to
1 ppm
1 to
20 ppm
2 to
30 ppm
30 to
60 ppm
Number of Pump Strokes
8 4 2 1
Correction Factor
1/4 1/2 1 2
Sampling Time 1 minute per pump stroke
Detecting Limit 0.1 ppm (n=8)
Colour Change Blue → Yellow
Reaction Principle SO2 + BaCl2 +H2O BaSO3 + 2HCl
HCl+Base Chloride
Coefficient of Variation 10% (for 2 to 10 ppm), 5% (for 10 to 30 ppm)
Shelf Life 3 Years
Corrections for temperature & humidity Unnecessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Carbon monoxide,
Nitric oxide
  No No
Carbon dioxide 100% Unclear demarcation Unclear demarcation
Nitrogen dioxide 1/1 + Pale purple
Hydrogen sulphide   No No
Other substance measurable with this detector tube
Substance Correction Factor No. of pump strokes Measuring Range
Thionyl chloride 0.72 2 1.44 to 21.6 ppm
Calibration gas generation Permeation tube method
TLV-TWA TLV-STEL Explosive range
2 ppm 5 ppm -