Water vapour |
NO.6L |
Measuring Range |
|
||
Number of Pump Strokes |
|
||
Correction Factor |
|
||
Sampling Time | 30 seconds per pump stroke | ||
Detecting Limit | 0.03 mg/L (n=1) | ||
Colour Change | Yellow → Purple | ||
Reaction Principle | H2O + Mg(ClO4)2 → Mg(ClO4)2 H2O | ||
Coefficient of Variation | 10% (for 0.05 to 0.2 mg/L), 5% (for 0.2 to 1 mg/L) | ||
Shelf Life | 3 Years | ||
Corrections for temperature & humidity | Unnecessary | ||
Store the tubes at cool and dark place. |
Substance | Concentration | Interference | Change colour by itself to |
Alcohols | Unclear demarcation | Purple (whole layer) | |
Aldehydes | Unclear demarcation | Purple (whole layer) | |
Ketones | Unclear demarcation | Purple (whole layer) | |
Nitriles | Unclear demarcation | Purple (whole layer) | |
Hydrocarbons | No | No | |
Nitrogen dioxide | No | No | |
Sulphur dioxide | 100 ppm | + | No |
Ammonia | 10 ppm | Unclear demarcation | Yellowish green |
Calibration gas generation | Diffusion tube method |
TLV-TWA | TLV-STEL | Explosive range |
- | - | - |