Water vapour |
NO.6LLP |
Measuring Range | 2 to 10 LB/MMCF |
Number of Pump Strokes | 2 |
Correction Factor | 1 |
Sampling Time | 3 minutes per pump stroke |
Detecting Limit | 2 LB/MMCF (n=2) |
Colour Change | Yellow → Green |
Reaction Principle | H2O + Mg(ClO4)2 → Mg(ClO4)2 H2O |
Coefficient of Variation | 10% (for 2 to 4 LB/MMCF), 5% (for 4 to 10 LB/MMCF) |
Shelf Life | 3 Years |
Corrections for temperature & humidity | Temperature correction is necessary |
Store the tubes at cool and dark place. |
Substance | Concentration | Interference | Change colour by itself to |
Sulphur dioxide | 15ppm | No | No |
Nitrogen dioxide,Hydrogen sulphide | 30ppm | No | No |
Hydrogen cyanide | 40ppm | No | No |
Methyl ethyl ketone | 70ppm | No | No |
Ethyl acetate | 100ppm | No | No |
Acrylonitrile | 40ppm | No | No |
Acetaldehyde | 50 ppm | No | No |
Methanol | 50 ppm | + | Green |
Triethylene glycol | No | No |
Calibration gas generation | Diffusion tube method |
TLV-TWA | TLV-STEL | Explosive range |
- | - | - |