Water vapour |
NO.6LP |
Measuring Range |
|
||
Number of Pump Strokes |
|
||
Correction Factor |
|
||
Sampling Time | 1.5 minutes per pump stroke | ||
Detecting Limit | 2 LB/MMCF (n=1) | ||
Colour Change | Yellow → Green | ||
Reaction Principle | H2O + Mg(ClO4)2 → Mg(ClO4)2 H2O | ||
Coefficient of Variation | 10% (for 3 to 10 LB/MMCF), 5% (for 10 to 40 LB/MMCF) | ||
Shelf Life | 3 Years | ||
Corrections for temperature & humidity | Temperature correction is necessary | ||
Store the tubes at cool and dark place. | |||
*(may become purple with high humidity) |
Substance | Concentration | Interference | Change colour by itself to |
Sulphur dioxide | 15ppm | No | No |
Nitrogen dioxide,Hydrogen sulphide | 30ppm | No | No |
Hydrogen cyanide | 45ppm | No | No |
Methyl ethyl ketone | 70ppm | No | No |
Ethyl acetate | 100ppm | No | No |
Acrylonitrile | 55ppm | No | No |
Acetaldehyde | 50 ppm | No | No |
Methanol | 100 ppm | + | Green |
Triethylene glycol | No | No |
Calibration gas generation | Diffusion tube method |
TLV-TWA | TLV-STEL | Explosive range |
- | - | - |