Phosphine |
NO.7 |
Measuring Range |
|
|||
Number of Pump Strokes |
|
|||
Correction Factor |
|
|||
Sampling Time | 1.5 minutes per pump stroke | |||
Detecting Limit | 0.5 ppm (n = 4) | |||
Colour Change | White → Pale yellow | |||
Reaction Principle | 2PH3 + 6HgCl2 + 3H2O → Hg3P2 3HgCl23H2O+6HCl | |||
Coefficient of Variation | 10% (for 5 to 10 ppm), 5% (for 10 to 50 ppm) | |||
Shelf Life | 3 Years | |||
Corrections for temperature & humidity | Unnecessary | |||
Store the tubes at cool and dark place. |
Substance | Concentration | Interference | Change colour by itself to |
Arsine | 20 ppm | + | Brown |
Hydrogen selenide | 20 ppm | + | Brown |
Hydrogen sulphide | 10 ppm | + | Dark brown |
Calibration gas generation | High pressure gas cylinder method |
TLV-TWA | TLV-STEL | Explosive range |
0.3 ppm | 1 ppm | - |