R-SH |
|
Performance
Measuring Range |
0.5 to 5 ppm |
5 to 120 ppm |
|
Number of Pump Strokes |
|
Correction Factor |
|
Sampling Time |
2 minutes per pump stroke |
Detecting Limit |
0.2 ppm (n=10) |
Colour Change |
White → Yellow |
Reaction Principle |
2RSH+ PdSO4 → (RS)2Pd+ H2SO4 |
Coefficient of Variation |
10% (for 5 to 40 ppm), 5% (for 40 to 120 ppm) |
Shelf Life |
3 Years |
Corrections for temperature & humidity |
Temperature correction is necessary |
Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences
(NOTE)
Substance |
Concentration |
Interference |
Change colour by itself to |
Carbon monoxide |
2000 ppm |
+ |
Pale brown (whole layer) |
Hydrogen sulphide |
500 ppm |
+ |
No |
Acetylene, Ethylene |
2000 ppm |
+ |
Pale brown (whole layer) |
Other substance measurable with this detector tube
Substance |
Correction Factor |
No. of pump strokes |
Measuring Range |
Ethyl mercaptan |
1.0 |
1 to 10 |
0.5 to 120 ppm |
Isopropyl mercaptan |
2.0 |
1 |
10 to 240 ppm |
Methyl mercaptan |
0.7 |
1 to 10 |
0.35 to 84 ppm |
Propyl mercaptan |
4.5 |
1 |
22.5 to 540 ppm |
Calibration gas generation |
Permeation tube method |
TLV-TWA |
TLV-STEL |
Explosive range |
- |
- |
- |