Mercaptans

NO.70L
RESH
Performance
Measuring Range
0.1 to 0.2 ppm 0.2 to 0.5 ppm 0.5 to 4 ppm 4 to 8 ppm
Number of Pump Strokes
4 2 1 1/2
Correction Factor
0.2 0.4 1 2
Sampling Time 1.5 minutes per pump stroke
Detecting Limit 0.05 ppm (n=4)
Colour Change Yellow → Red
Reaction Principle RSH + HgCl2 → RSHgCl + HCl
HCl + Base ¨ Chloride
Coefficient of Variation 10% (for 0.5 to 1 ppm), 5% (for 1 to 4 ppm)
Shelf Life 2 Years
Corrections for temperature & humidity Unnecessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Acid gases   + Red
Hydrogen sulphide 200 ppm No No
Other substance measurable with this detector tube
Substance Correction Factor No. of pump strokes Measuring Range
Butyl mercaptan 1.6 1/2,1,2 or 4 0.16 to 12.8 ppm
tert-Butyl mercaptan 1.0 0.1 to 8 ppm
Ethyl mercaptan 1.0 0.1 to 8 ppm
Methyl mercaptan 1.0 0.1 to 8 ppm
Propyl mercaptan 1.2 0.12 to 9.6 ppm
Calibration gas generation Permeation tube method
TLV-TWA TLV-STEL Explosive range
- - -