Ethyl mercaptan |
NO.72 |
Measuring Range |
|
||
Number of Pump Strokes |
|
||
Correction Factor |
|
||
Sampling Time | 2 minutes per pump stroke | ||
Detecting Limit | 0.2 ppm (n=10) | ||
Colour Change | White → Yellow | ||
Reaction Principle | 2C2H5SH + PdSO4 → (C2H5S)2Pd+ H2SO4 | ||
Coefficient of Variation | 10% (for 5 to 40 ppm), 5% (for 40 to 120 ppm) | ||
Shelf Life | 3 Years | ||
Corrections for temperature & humidity | Temperature correction is necessary | ||
Store the tubes at cool and dark place. |
Substance | Concentration | Interference | Change colour by itself to |
Acetylene, Ethylene | 2000 ppm | + | Pale brown (whole layer) |
Carbon Monoxide | 2000 ppm | + | Pale brown (whole layer) |
Other Mercaptans | + | Yellow | |
Hydrogen sulphide | 500 ppm | No | No |
Calibration gas generation | Diffusion tube method |
TLV-TWA | TLV-STEL | Explosive range |
0.5 ppm | - | 2.8 to 18.2% |