Tetrahydrothiophene |
NO.76 |
Measuring Range | 1 to 10 ppm |
Number of Pump Strokes | 4 |
Correction Factor | 1 |
Sampling Time | 1.5 minutes per pump stroke |
Detecting Limit | 0.25 ppm (n=4) |
Colour Change | Pink → Pale Yellow |
Reaction Principle | C4H8S + MnO4 (Pink) → Pale Yellow Product |
Coefficient of Variation | 10% (for 1 to 3 ppm), 5% (for 3 to 10 ppm) |
Shelf Life | 2 Years |
Corrections for temperature & humidity | Temperature correction is necessary |
Store the tubes at cool and dark place. |
Substance | Concentration | Interference | Change colour by itself to |
Ethyl mercaptan | 1 ppm | No | Pale pink (whole layer) |
Ethylene | 1000 ppm | No | Pale pink (whole layer) |
Propylene | 500 ppm | No | Pale pink (whole layer) |
Methanol | 300 ppm | No | Pale pink (whole layer) |
Hydrogen sulphide | 30 ppm | No | Pale pink (whole layer) |
Calibration gas generation | Diffusion tube method |
TLV-TWA | TLV-STEL | Explosive range |
- | - | - |