Detector Tube |
|
Measuring Range |
1 to 15 mg/m3 |
1 to 15 mg/m3 |
|
Number of Pump Strokes |
|
Correction Factor |
|
Sampling Time |
3 minutes per pump stroke |
Detecting Limit |
0.2 mg/m3 (n=1) |
Colour Change |
Yellow → Pink |
Pink → Pale yellow |
|
Reaction Principle |
(CH3)3CSH + HgCl2 → (CH3)CSHgCl + HCl
HCl + Base → Chloride compound
(CH3)2S + KMnO4 (Pink) → Pale yellow product |
Coefficient of Variation |
10% (for 1 to 5 mg/m3), 5% (for 5 to 15 mg/m3) |
Shelf Life |
2 Years |
Corrections for temperature & humidity |
Temperature correction is necessary |
Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below. |