Cl2 |
|
Performance
Measuring Range |
0.1 to 0.5 ppm |
0.5 to 8 ppm |
8 to 16 ppm |
|
Number of Pump Strokes |
|
Correction Factor |
|
Sampling Time |
1 minutes per pump stroke |
Detecting Limit |
0.05 ppm (n=5) |
Colour Change |
White → Pale pink |
Reaction Principle |
Cl2 + o - Tolidine → Pale pink products |
Coefficient of Variation |
10% (for 0.5 to 2 ppm), 5% (for 2 to 8 ppm) |
Shelf Life |
3 Years |
Corrections for temperature & humidity |
Unnecessary |
Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences
(NOTE)
Substance |
Concentration |
Interference |
Change colour by itself to |
Nitric oxide |
|
+ |
Pale pink |
Ozone |
|
+ |
Pale pink |
Nitrogen dioxide |
|
+ |
Pale pink |
Chlorine dioxide |
|
+ |
Pale pink |
Bromine, Iodine |
|
+ |
Pale pinky |
Other substance measurable with this detector tube
Substance |
Correction Factor |
No. of pump strokes |
Measuring Range |
Bromine |
0.1 |
4 |
0.05 to 0.8 ppm |
Calibration gas generation |
Permeation tube method |
TLV-TWA |
TLV-STEL |
Explosive range |
0.5 ppm |
1 ppm |
- |