Formaldehyde

NO.91
HCHO
Performance
Measuring Range
2 to 20 ppm 20 to 50 ppm 50 to 100 ppm
Number of Pump Strokes
2 1 1/2
Correction Factor
1 2.5 5
Sampling Time 1 minute per pump stroke
Detecting Limit 0.5 ppm (n=2)
Colour Change White → Brown
Reaction Principle HCHO + C6H4(CH3)2 + H2S2O7 → Condensation Polymer
Coefficient of Variation 15% (for 2 to 6 ppm), 10% (for 6 to 20 ppm)
Shelf Life 3 Years
Corrections for temperature & humidity Unnecessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Aldehydes 1/2 time + Brown
Esters, Ethers 1000 ppm + No
Ketones 1000 ppm + No
Styrene 500 ppm + Yellow
Calibration gas generation Diffusion tube method
TLV-TWA TLV-STEL Explosive range
- C 0.3 ppm 7.0 to 73%