Formaldehyde |
NO.91 |
Measuring Range |
|
|||
Number of Pump Strokes |
|
|||
Correction Factor |
|
|||
Sampling Time | 1 minute per pump stroke | |||
Detecting Limit | 0.5 ppm (n=2) | |||
Colour Change | White → Brown | |||
Reaction Principle | HCHO + C6H4(CH3)2 + H2S2O7 → Condensation Polymer | |||
Coefficient of Variation | 15% (for 2 to 6 ppm), 10% (for 6 to 20 ppm) | |||
Shelf Life | 3 Years | |||
Corrections for temperature & humidity | Unnecessary | |||
Store the tubes at cool and dark place. |
Substance | Concentration | Interference | Change colour by itself to |
Aldehydes | 1/2 time | + | Brown |
Esters, Ethers | 1000 ppm | + | No |
Ketones | 1000 ppm | + | No |
Styrene | 500 ppm | + | Yellow |
Calibration gas generation | Diffusion tube method |
TLV-TWA | TLV-STEL | Explosive range |
- | C 0.3 ppm | 7.0 to 73% |