NO2 |
|
Performance
Measuring Range |
0.5 to 30 ppm |
30 to 125 ppm |
|
Number of Pump Strokes |
|
Correction Factor |
|
Sampling Time |
1 minute per pump stroke |
Detecting Limit |
0.1 ppm (n=2) |
Colour Change |
White → Yellowish Orange |
Reaction Principle |
NO2 + o-Tolidine Yellowish orange product |
Coefficient of Variation |
10% (for 0.5 to 3 ppm), 5% (for 3 to 30 ppm) |
Shelf Life |
3 Years |
Corrections for temperature & humidity |
Unnecessary |
Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences
(NOTE)
Substance |
Concentration |
Interference |
Change colour by itself to |
Ammonia |
|
No |
No |
Carbon monoxide |
|
No |
No |
Carbon dioxide |
|
No |
No |
Nitric oxide |
50 ppm |
Unclear demarcation |
Pale red |
Bromine, Chlorine |
1/5 times |
+ |
Yellowish orange |
sulphur dioxide |
10 ppm |
(Bleaching) |
No |
Organic solvents |
|
No |
No |
Other substance measurable with this detector tube
Substance |
Correction Factor |
No. of pump strokes |
Measuring Range |
Iodine |
0.4 |
2 |
0.2 to 12 ppm |
Calibration gas generation |
Permeation tube method |
TLV-TWA |
TLV-STEL |
Explosive range |
3 ppm |
5 ppm |
- |