Explanation of specification items

(Example)

Styrene

NO.124L
HCHO
Performance
Measuring Range 2 to 25 ppm 25 to 100 ppm
Number of Pump Strokes 4 1
Correction Factor 1 4
Sampling Hours 30 seconds per pump stroke
Detecting Limit 0.5 ppm (n=4)
Colour Change White → Yellow
Reaction Principle Styrene reacts with sulphuric acid fuming to form a condensation polymer which is yellow stain.
Coefficient of Variation 10% (for 2 to 5 ppm), 5% (for 5 to 25 ppm)
Shelf Life 3 Years
Corrections for temperature & humidity Unnecessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences
Substance Concentration Interference Change colour by itself to
Alcohols, Esters 10 times + (Bleaching) No
Aldehydes, Ketones 10 times + (Bleaching) No
Butadiene 5 ppm + (Bleaching) Blackish brown
Other substance measurable with this detector tube
Substance Correction Factor Pump Strokes Measuring Range
Divinyl benzene 0.6 3 1 to 15 ppm
Calibration gas generation Diffusion tube method
TLV-TWA TLV-STEL Explosive range
20 ppm 40 ppm 1.1 to 6.1%


Return to previous viewGo to top of page