Single Gas Detector Docking Station
The ARA DOCK4 by ION Science is an automatic test and calibration station for ARA single gas detectors. With one button, it runs bump tests, calibration, buzzer and LED tests and data log transfers on up to four detectors at once. It records results with the serial number, date and time for compliance, and an internal battery gives about 5 hours of operation away from mains power.
KEY BENEFITS
- Tests up to 4 ARA single gas detectors at once, cutting testing time
- Records calibration and bump test results with serial number, date and time for compliance documentation
- One button operation for bump tests, calibration, buzzer and LED tests and data log transfers
- Compatible with ARA 100, 100H, 200, 200H, 300 and 400 detectors
- Built in calibration gas cylinder holder with pressure gauge, so remaining gas is easy to check
+ more in the Description tab below
Four ARA Detectors. One Button.
For sites running several ARA single gas detectors, testing them one by one takes time. The ARA DOCK4 tests up to four ARA detectors at once, with a single button press.
It handles the essential maintenance tasks: bump testing, calibration, buzzer testing, LED testing and data log transfers. Each result is recorded with the detector’s serial number, date and time, creating a compliance record without manual data entry.
Compatible ARA Detectors
The DOCK4 works with ARA 100, 100H, 200, 200H, 300 and 400 single gas detectors, covering H2S, CO, O2 and SO2. Detectors are recognised automatically through IrDA communication.
The DOCK4 uses one calibration gas cylinder, held in the dock with a pressure gauge to show remaining gas. To test detectors for different gases in the same session, use a cylinder that contains all the gases they need. ION recommends span gas concentrations of 25 ppm H2S, 100 ppm CO, 18% O2 and 10 ppm SO2, and these can be adjusted.
Simple to Run, Easy to Manage
Unit LEDs show the status of each detector during testing. The DOCK4 runs on a 12 V mains adapter, and its internal rechargeable battery gives about 5 hours of continuous operation with four detectors docked.
Event logs are stored on an internal 16 GB USB memory stick. Connected to a PC, the Dock4Manager software lets authorised users change test settings and span gas values, read event and test logs, and upgrade firmware.
The DOCK4 has a full 2 year warranty from the date of shipment. It’s designed for use in non hazardous locations only.
Benefits
- Tests up to 4 ARA single gas detectors at once, cutting testing time
- Records calibration and bump test results with serial number, date and time for compliance documentation
- One button operation for bump tests, calibration, buzzer and LED tests and data log transfers
- Compatible with ARA 100, 100H, 200, 200H, 300 and 400 detectors
- Built in calibration gas cylinder holder with pressure gauge, so remaining gas is easy to check
- For use in non-hazardous locations only
- Internal battery gives about 5 hours of continuous operation with 4 detectors docked
- Dock4Manager PC software to change test settings, span gas values and firmware
- Event logs stored on an internal 16 GB USB memory stick
- Full 2 year warranty from date of shipment
Compatible instrument
ARA 100, 100H, 200, 200H, 300 and 400
Capacity
Up to 4 detectors at once
Tests
Bump, calibration, data transfer, audible alarm, visual alarm
Recommended span gases
H2S 25 ppm, CO 100 ppm, O2 18%, SO2 10 ppm (adjustable)
Gas cylinder
Built in holder. Compatible with 116 L cylinders (C10 valve) and 2AL, 8AL, 34L and 58L types. Pressure gauge up to 1,500 psi
Gas in/out
1 purge (ambient air) inlet and 1 exhaust outlet, 1/8 inch hose barb
Pumps
2 built in, 250 to 350 mL/min
Detector recognition
IrDA communication
Display
4 two colour unit LEDs. Power, Bump and Cal button LEDs
Data storage
Internal 16 GB USB memory stick for event logs. Time and date stamp
Interface
USB 2.0 (type B), RJ45 Ethernet 10/100
Software
Dock4Manager PC software
Power supply
110 to 240 V AC input, 12 V DC 1.5 A output. Internal Li-ion battery 7.4 V, 2,750 mAh
Battery runtime
About 5 hours continuous with 4 detectors docked
Dimensions
496.8 x 340 x 115 mm
Weight
7 kg (without gas cylinder)
Enclosure
Aluminium
Operating temperature
5°C to 40°C
Humidity
10 to 90% RH non condensing
Safety rating
For use in non hazardous locations only
Certifications
CE, FCC, UKCA
Warranty
Full 2 years from date of shipment
- Aviation & Transportation
- Chemical Manufacturing
- Confined Space & Compliance
- Energy (Oil & Gas)
- Food & Beverage
- Landfill & Remediation
- Marine & Maritime
- Medical & Healthcare
- Mining & Minerals
- Power Generation & BESS
- Technology & Semiconductor
- Water & Wastewater
The ARA DOCK4 tests up to four ARA single gas detectors simultaneously in a single session.
The DOCK4 works with ARA 100, 100H, 200, 200H, 300 and 400 single gas detectors. It isn't compatible with the ARA-X4 multi gas detector. For the ARA-X4, use the ARA-X Docking Station.
Testing up to four detectors in one session cuts testing time compared with testing them one by one. Gas use depends on the test settings and the cylinder used. Talk to our Technical Consultants about the right cylinder for your fleet.
No. The ARA DOCK4 is rated for use in non-hazardous locations only.